MR Imaging Protocol
Sequence |
TE | TR | Flip | FOV | Matrix | Slice Thickness (mm) | Slice Gap (mm) | b-value | NEX | Options | Acquisition time (min:sec) |
Sagital T1 localizer | 14 | 500 | 200X400 | 192X256 | 5 | 1.5 | 1 | 1:37 | |||
Axial FSE T2 | 102 | 6000 | 200X400 | (.75) 256X320 | 5 | 1 | 2 | 3:13 | |||
Axial FLAIR | 120 | 10,000 | TI=2200 | 200X00 | 192X256 | 5 | 1 | 1 | 4:40 | ||
Coronal oblique T2 FSE | 102 | 6000 | 200X200 | (.75) 224X320 | 4 | 1 | 3 | 3:42 | |||
Coronal oblique FLAIR | 120 | 10,000 | TI=2200 | 200X200 | 192X256 | 4 | 1 | 1 | 4:40 | ||
Coronal SPGR volumetric dataset | 8 | 30 | 35 | 200X200 | 160X256 | 1.5 | 0 | 1 | 10:18 | ||
Axial DWI | 72.9 | 7500 | 400X200 | 256X128 | 5 | 1 | 1000 s/mm2 | 1 | EPI | 2:38 | |
Axial Susceptibility GRE (Optional) | 25 | 750 | 20 | 200X200 | (.75) 192X256 | 5 | 1 | 2 | 3:41 | ||
Single Voxel Spectroscopy | 144 | 1500 | 200X400 | 8 |
Abbreviations:
DWI | = Diffusion Weighted Imaging |
FLAIR | = Fluid Attenuated Inversion Recovery |
Flip | = Flip angle for gradient echo |
FOV | = Field of View |
FSE | = Fast Spin Echo |
GRE | = Gradient Echo |
Matrix | = Number of pixels in x and y direction |
SPGR | = Spoiled gradient recalled echo |
TE | = Echo Time |
TI | = Inversion time |
TR | = Repetition TimeTE |
Note: These parameters are customized for use on scanners at MGH but may not be optimal at other sites. No assurances are expressed or implied.